Initiate Subsystem Commands - Keysight N774 C Series Programming Manual

Multiport power meter
Table of Contents

Advertisement

4
Measurement Operations & Settings

INITiate Subsystem Commands

Command:
:INITiate[n][:CHANnel[m]][:IMMediate]
Syntax:
:INITiate[n][:CHANnel[m]][:IMMediate]
Description:
Initiates the software trigger system and completes one full trigger cycle, that is, one measurement is made for selected [n].
In logging mode it triggers all channels independent from [n].
Parameters:
None
Response:
None
Example:
init1:imm
init2:imm
Affects:
N7744C and N7745C
Command:
:INITiate[n][:CHANnel[m]]:CONTinuous
Syntax:
:INITiate[n][:CHANnel[m]]:CONTinuous<wsp><boolean>
Description:
Sets the software trigger system to continuous measurement mode.
Parameters:
A boolean value:
0 or OFF: do not measure continuously
1 or ON: measure continuously
Response:
None
Example:
init2:cont 1
Affects:
N7744C and N7745C
66
This section provides the description of the following commands.
Keysight N774xC Series Multiport Power Meter Programming Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7744cN7745c

Table of Contents