Set Up Software for Linear Measurement

1
Figure 9. Metrology Main Display
Keysight 5519A and 5519B User's Guide
Place the interferometer assembly about 1 cm (1/2 in) away from the
straightedge approximately 5 cm (2 in) from the front of the laser head as
shown in
Figure 5
.
Place the retroreflector against the straightedge as shown in Figure 5.
Place the retroreflector as close as possible to the interferometer assembly
without allowing the two to touch.
Mark the positions of the interferometer assembly and retroreflector on the
template with a pencil or piece of tape. You will later move the optics and will
have to return them to the position you marked.
You are now ready to set up the metrology software.
Start the Metrology Software; the Main Display is shown in Figure 9.