Overhead Time; To Control Source/Measure Timing (B2985B/B2987B) - Keysight B2980B Series User Manual

Femto/picoammeter electrometer/high resistance meter
Table of Contents

Advertisement

Function Details
Measurement Time

Overhead Time

Overhead time is the time required for range change, etc. This time depends on
the measurement condition, and cannot be specified. A major element of the
overhead time is range change time during measurement (when measurement
ranging mode is set to AUTO).

To Control Source/Measure Timing (B2985B/B2987B)

The voltage source output and measurement timing can be controlled by the
following parameters. See
output. For a bias output, focus the shape of a sweep step only.
1. Source delay
2. Measure delay
3. Period
208
The source delay time is defined as the time from trigger to start of a voltage
source output.
The measurement delay time is defined as the time from trigger to start of a
measurement.
The period is the trigger interval. It can be defined for the TIMER or MANUAL
trigger type (trigger source) and for the voltage source output and
measurement actions individually.
Figure
5-2. This figure shows an example of sweep
Keysight B2980B User's Guide, Edition 1