System Messages; Status Displays; Error Messages; Ex Yy Zz - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

2.2 System messages

System code
Description
S100
ORTHOPHOS XG is started
S110
Exposure cannot be performed.

Status displays

Plus
ready
(only for XG
/ 3D
Ready for exposure
Please wait
Ready for exposure in XX seconds
Exposure in progress
Error code: Ex yy zz
2 – 6
2.2

System messages

2.3

Status displays

Description
)
The unit is ready for operation.
The unit waiting for operational readiness.
The cooling time countdown is running.
Radiation is released.
2.4

Error messages

Error messages are displayed as error codes (Ex yy zz) on the touchscreen
of the Easypad (XG
PPE) or on the remote control display (if available).
The codes provide you with error type, error location and troubleshooting
information.
The error messages are encoded according to the following pattern:
Error type
"Troubleshooting"
classification for the
The error messages are sorted by modules in the table on page 2-7.
Actions required
Wait, no action required
Quit readiness for exposure,
switch unit OFF, wait for 1 minute, switch
unit back ON and observe error messages
displayed after switch-on.
Repeat procedure if necessary
Plus
ready
/ 3D
), on the display of the Multipad (XG5 / 3 / 3

Ex yy zz

Location
Module, subsystem or logi-
cal function unit
Tab 2
Sequential number
with identification
of the error
59 38 399 D3352
D3352.076.01.26.02 09.2016