7 Malfunctions
7.3 Error description
64
E6 – Self-check
The error occurs spontaneously and without a corresponding operation.
1. Acknowledge the error message.
The error no longer occurs.
2. If the error remains, switch the unit off and on again. NOTICE!
Observe waiting period!
The error no longer occurs.
3. If the error persists, contact your Customer Service.
E7 – Serious system error
The error occurs spontaneously and without a corresponding operation.
1. Switch off the unit.
2. Contact your Customer Service immediately.
The unit is functional.
7.3.2
yy – Locality
The location may be a DX module number standing for an entire HW
function unit, or a logical SW function unit on board DX11 (central
control).
10 – Central control DX 11; system hardware
11 – Central control DX 11; system software
12 – Central control DX 11; central CAN bus fault
13 – Central control DX 11; peripheral DX11, DX1 (tripod mechanics,
tripod sensors)
14 – Central control DX 11; digital extension (HSI, network, etc.)
15 – Central control DX 11; configuration (wrong software, wrong module
constellation, etc.)
06 – X-ray tube assembly
71 – Multipad user interface
DX71 – Multipad
42 – Remote
61 – Diaphragm control
Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG 3
60 50 681 D3352
D3352.201.05.18.02 04.2016