Yy - Location; General Handling Of Error Messages - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
yy
Location/Function unit
06
X-ray tube assembly
61
Diaphragm control
07
Easypad user interface
10
System hardware
11
System software
12
CAN bus
13
Stand peripherals
14
Digital extension
15
Configuration/update (wrong software, wrong module constellation, etc...)
42
Remote control
81
2D sensor (PAN) or Ceph sensor
88
Combisensor
91
Cephalometer
63 03 510 D3352
D3352.076.03.11.02 09.2016
5.4.3

yy - Location

Identifier yy defines the location or logical function unit where the error
has occurred.
The location may be a DX module number standing for an entire HW
function unit, or a logical SW function unit on board DX11 (central
control).
5.4.4

General handling of error messages

Error messages must always be acknowledged with the R key.
If trouble-free operation is possible after the error is acknowledged, then
no further action is necessary.
If error messages occur again or frequently, or if fault-free operation is not
possible, run an error diagnosis (see chapter "Troubleshooting [ → 160]").
In some cases, it may make sense to obtain more information on the
history and frequency of errors via the error logging memory (S007) and
"SiXABCon"
via
[ → 71]) (see also section "Error logging memory [ → 161]").
"Properties"
"Details..."
/
/
(see section, Opening "Details"
5 Messages
5.4 Error messages
Board
DX6
DX61
DX7
DX11/DX1
DX11/DX1
DX11/DX1
DX11/DX1
DX11/DX1
DX11/DX1
DX42
DX81
DX88
DX91
93