7 Malfunctions
7.3 Error description
92
E6 – Self-check
The error occurs spontaneously and without a corresponding operation.
1. Acknowledge the error message.
The error no longer occurs.
2. If the error remains, switch the unit off and on again. NOTICE!
Observe waiting period!
The error no longer occurs.
3. If the error persists, contact your Customer Service.
E7 – Serious system error
The error occurs spontaneously and without a corresponding operation.
1. Switch off the unit.
2. Contact your Customer Service immediately.
The unit is functional.
7.3.2
yy – Locality
The location may be a DX module number standing for an entire HW
function unit, or a logical SW function unit on board DX11 (central
control).
06 – Tube assembly
71 – Multipad user interface
10 – Central control DX 11; system hardware
11 – Central control DX 11; system software
12 – Central control DX 11; central CAN bus fault
13 – Central control DX 11; DX11, DX1 periphery (motor system of stand,
sensor system of stand)
14 – Central control DX 11; digital extension (HSI, network, etc.)
15 – Central control DX 11; configuration (wrong software, wrong module
constellation, etc.)
16 – Central control DX 11; Zero Management
20 – Central control DX 11; Framegrabber application
22 – Central control DX 11; 2D Imaging System (Ajat, FP)
23 – Central control DX 11; 3D Imaging System (FP)
42 – Remote
61 – Diaphragm control
81 – Sensor
91 – Ceph digital
Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG 5 / Ceph
D3352.201.04.25.23 12.2017
60 04 811 D3352