7 Installation: Remote control
7.4 Connecting the control cables (REMOTE)
A
68
7.4

Connecting the control cables (REMOTE)

7.4.1

Installation version 1: Without coiled cable

X108
XRAY
XRAY
GND
DX42
+28 V
CAN-H
CAN-L
PIN
1
2
3
4
5
6
1. Unscrew clamp (A) from the board.
2. Place cable L117 in the clamp so that the turned up cable shield is
completely enclosed.
3. Re-attach the clamp to the board.
4. Connect control cable L117 to terminal X108 (board DX42) as shown
in the connection diagram. Shorten control cable L117 as required.
5. Lay cable L117 in a loop following the bottom edge of the chassis.
6. Close the housing.
For information on how to connect a door contact switch, refer to the
section entitled "Connecting a door contact switch [ → 72]".
Installation Manual ORTHOPHOS XG 3D / Ceph
X108
Pin
1
1
2
2
3
3
4
4
5
5
6
6
Code
WHGN
GN
BN
WHBN
BU/WHOG
WHBU/OG
Sirona Dental Systems GmbH
15 m
590"
L117
L117
WHGN
GN
BN
WHBN
BU/WHOG
WHBU/OG
Appearance
White/green
Green
Brown
White/brown
Blue / white/orange
White/blue / orange
63 03 452 D3352
D3352.031.05.08.02 10.2015