Replacing Cables; Replacing Fiber-Optic Cable L5, L6 Or L15 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016
9.17.2

Replacing cables

CAUTION
Switch the unit off before you start replacing cables or removing
connectors.
NOTICE
Be careful not to twist the cables or kink the fiber-optic light guides when
installing them.
Always check the cables before replacing them [ → 176].
Cable replacement; XG 3D
The cables are labeled with small flags. They specify the designation and
part number of the cable. The plugs and sockets on the cables are
designated both on the boards and cables. Check the designation when
you pull off the cables.
Some cables feature markings of green adhesive tape. Mark the
corresponding positions on the unit before removing an old cable. Lay the
new cable so that the cable markings again come to rest at the
corresponding positions marked on the unit while removing the old cable.
An overview of all cables can be found in the chapter Overview of cabling
[ → 26].
9.17.2.1

Replacing fiber-optic cable L5, L6 or L15

1. Remove the defective fiber-optic cable.
2. NOTICE! Do not kink or twist fiber-optic cables, the bending radius
may not be less than 20mm, otherwise it is at risk of breaking!
Clip the two parts of the first radius limiter (A) close to the connector,
which is plugged onto board DX1, onto the cable.
3. Plug the connector of the new fiber-optic cable to the same color
assignment on the board DX1.