What Has To Be Done After Replacing The Light Barriers - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
V61_4
After replacing V1_1, V1_2 or V1_3
After replacing V1_4
After replacing V61_4 or V61_5
After replacing V91_1 or V91_2
63 03 510 D3352
D3352.076.03.11.02 09.2016
● Light barrier at combisensor (rotation unit), 3D/Ceph position: V61_4
● Light barrier at combisensor (rotation unit), 2D position: V61_5
V61_5
Cephalometer
● Light barrier for patient diaphragm position detection, ceph: V91_1
● Light barrier for sensor position detection, ceph: V91_2
9.15.2

What has to be done after replacing the light barriers?

➢ Perform a complete unit adjustment or calibration.
➢ Perform a function check.
➢ Perform a complete unit adjustment or calibration.
1. Perform service routines S034.4 [ → 348] and S034.5 [ → 352].
2. Check the ceph adjustment using the test exposure Ceph - Fixed
point of rotation test exposure.
3. Perform a ceph unit adjustment, if necessary.
V91_2
9 Repair
9.15 Light barriers
V91_1
427