8 Perform service routines via the control panel
8.20 S034: Service for the digital cephalometer
SR*
S034
S034.4
S034.5
S034.6
S034.8
S034.9
348
8.20

S034: Service for the digital cephalometer

SHZ**
Function
Service for the digital cephalometer
No
Display/calibrate center position of Ceph scan sensor axis
No
Display/calibrate center position of Ceph scan secondary diaphragm
axis
No
Moving to the Ceph packing position
No
User-specific determination of the shadowing limit for C1 and C2
No
User-specific determination of the shadowing limit for C3 and C3F
* SR=service routine, ** SHZ=security access
8.20.1

S034: Test step 4

Display/calibrate center position of Ceph scan sensor axis
1. Detach the cover to the cephalometer [ → 41].
2. Call service routine S034.4 [ → 262].
3. Touch the T key (G).
4. Move the sensor to its front position by touching the R key (I). The
sensor can always be moved back and forth between its front and
center positions with the R key.
5. Move the sensor to its center position by touching the R key.
Service Manual ORTHOPHOS XG 3D / Ceph
After selecting the service routine, the current offset value for the
center position of the ceph scan
selection field 1.
The test mode is activated.
In the test mode, travel to the center position of the ceph scan
sensor axis is initially performed without offsets. The T key lights
up and indicates the test mode.
Sirona Dental Systems GmbH
sensor axis
is displayed in
63 03 510 D3352
D3352.076.03.11.02 09.2016