S032: Test Step 50; Possible Results Of Self-Test And Troubleshooting Measures - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

8 Perform service routines via the control panel
8.18 S032: Sensor test
"Success"
"ETimeout"
"Unknown"
"No Image"
344
8.18.2

S032: Test step 50

Self-test of 3D image data path
1. Call service routine S032.50.
"OP3D" is displayed in selection field 1.
2. Press the T key to start the DX88 self-test.
The self-test lasts approx. 10-15 seconds. A progress bar is
displayed as long as the self-test is active.
After the self-test is complete, the result of the test is shown in
selection field 1.
3. Analyze the result and take any troubleshooting measures, if
necessary (see Possible results of self-test and troubleshooting
measures [ → 344]).
4. Exit the service routine [ → 262].
8.18.2.1

Possible results of self-test and troubleshooting measures

The self-test was successfully executed. No errors were detected in the
image data path.
The self-test could not be completed within 20 seconds.
The self-test was canceled with an error that could not be localized.
An error occurred during the activation of the exposure for the self-test. If
no further error messages (see title bar of Easypad) occur, please
implement the following tests and measures:
● Check CAN bus [ → 163], specifically the CAN bus connection to
DX88
● Replace board DX88.
● Replacing board DX11
If error messages are present, please implement the measures according
to the displayed error message.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016