S021: Service For Motor-Driven Diaphragm; S021: Test Step 1 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
SR*
S021
S021.1
63 03 510 D3352
D3352.076.03.11.02 09.2016
8.17

S021: Service for motor-driven diaphragm

SHZ**
Function
Service for motor-driven diaphragm
No
Initialization of diaphragm axis
* SR=service routine, ** SHZ=security access
8.17.1

S021: Test step 1

Initialization of diaphragm axis
Selection field
1
2
1. Call service routine S021.1.
2. Touch the T key (C).
8 Perform service routines via the control panel
8.17 S021: Service for motor-driven diaphragm
Display
Inactive progress
indicator
Active progress
indicator
FFFF
0000
EEEE
An inactive progress indicator in selection field 1 and the
message "FFFF" in selection field 2 signal that the system is
ready for initialization of the diaphragm axis.
Automatic axis initialization of the diaphragm starts.
Meaning
Ready for initialization
Initialization is running
Ready for initialization
Initialization OK
Initialization faulty
339