S018: Test Step 2 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
*
63 03 510 D3352
D3352.076.03.11.02 09.2016
8.15.1

S018: Test step 2

Set the maximum travel height
1. Move the unit to the required maximum travel height by pressing the
S018_2 XG 3D / XG 3D ready
Up/Down keys in the user mode on the control panel.
2. Call service routine S018.2.
Once the service routine has been selected, the current height
position is displayed in selection field 1.
The Memory key (B) lights up.
3. To save the maximum travel height, touch the Memory key (B) (R key
(C) lights up) followed by the R key (C).
4. Set the mechanical limit stop at the unit:
Loosen nut (D) and move mechanical limit stop (E) for the limit switch
D
until it engages. Tighten the nut (D) again.
The next time the UP key is pressed, the unit stops 10 mm below the
limit switch.
5. Exit the service routine [ → 262].
E
8 Perform service routines via the control panel
8.15 S018: Service for height adjustment
325