S017: Test Step 2 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

8 Perform service routines via the control panel
8.14 S017: Configuration service
304
8.14.1

S017: Test step 2

Configuring the hardware version
S017_2 XG 3D
Selection field
1
* Factory setting
1. Call service routine S017.2.
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use the arrow keys (A) to select the code for the required hardware
version in selection field 1 (see table).
Once the hardware version has been selected, the Memory key
lights up.
3. Save the setting [ → 262].
4. Exit the service routine [ → 262].
Service Manual ORTHOPHOS XG 3D / Ceph
Code
Function
0001
Panoramic digital
0003
Panoramic digital, ceph left digital
0005
Panoramic digital, ceph right digital
0201
XG 3D + panoramic digital*
0203
XG 3D +
panoramic digital, ceph left digital
0205
XG 3D +
panoramic digital, ceph right digital
Sirona Dental Systems GmbH
63 03 510 D3352
D3352.076.03.11.02 09.2016