9 Adjusting / calibrating the system

9.1 System adjustment menu

9
Adjusting / calibrating the system
1.
Message
window
9.1
System adjustment menu
The
SERVICE FUNCTIONS
the calibration process. This service routine is started from
SIDEXIS XG:
UTILITIES
CONSTANCY TEST
)
DEVICE
SERVICE EXPOSURE
(
-
SELECT X
RAY COMPONENT
TIONS menu
-
The queries
SELECT X
RAY DEVICE
will only appear if several units are set up in
COMPONENT
SIDEXIS XG.
i
NOTE
The
SERVICE FUNCTIONS
password, enter the first four digits of the current system
date (PC) in reverse order.
Example: On 05/30/2004, the service password is 5003
i
NOTE
When you open the SERVICE FUNCTIONS menu, the unit
switches from the user mode to the PC service mode
logged by the PC.
For GALILEOS:
This mode is displayed on the Easypad via the PC service
image (2.).
For GALILEOS GAX5:
This mode is displayed on the Multipad via the "SERVICE"
display (3.).
.
18
menu tab
Tools pictogram
menu (1.) will guide you through
(
D
SELECT X
3
password query
)
SERVICE FUNC
-
and
SELECT X
RAY
menu is password-protected. As
2.
Preview
Image
Status column
In PC service mode, the control options that are available
on the control panel are determined by SIDEXIS XG and the
currently selected service routine. General control of the
unit by means of the control panel (as in the user mode) is
not possible in this mode.
-
RAY
The
SERVICE FUNCTIONS
-
that can be selected using the menu tabs.
The tools pictogram shows which (if any) test phantom
should be used for this particular adjustment step.
The message window displays text messages regarding
the adjustment process..
To the right of the menu you can see the status column.
This column contains information about the system's cur-
rent calibration state.
green
and
checked
green
red
To quit the
SERVICE FUNCTIONS
Sirona Dental Systems GmbH
Maintenance Instructions GALILEOS
3.
SERVICE
menu contains 9 sub-menus
valid data record,
calibration is in progress.
No calibration required!
calibration data record present, calibra-
tion has not been performed, calibration
might not be sufficient.
Calibration required!
invalid data record or no record present
Calibration required!
menu, click
61 16 813 D3437
D3437.103.01.04.02
.
CANCEL