3. Assure the lid is attached vertically on the bottom adapter, as shown.
4. Enter the "manual water channels operations" option by pressing the
"Left"-Key from the ready screen, following by the "Right"-Key.
5. Press the "Down"-Key.
 Water is pumped through the internal oil hoses and channels of
the lid.
6. When water is visible from underneath the lid's top plate; observe that
water is flowing freely from the 6 different adapters.
7. Repress the "Down"-Key.
8. Remove the white extension piece for oil from the chamber and the
lid.
9. Repeat the process for all available STANDARD lids.
10. Press the "Up"-Key to return to the ready screen.
 The water and oil residue that is in the chamber and hoses will need
to be drained and cleaned out, see section "Drain the chamber" in
"Drain the water tank via the menu [ → 125]".
6.5
Drain the water tank of the DAC UNIVERSAL
ADVANCED
Frequency: weekly or every 50 cycles (whichever comes first).
IMPORTANT
If you have installed the DAC UNIVERSAL with a permanent water
connection, please turn off the water.
You have the following options for emptying the water tank:
● Drain the water tank via the menu
● Drain the water tank via a hose
6.5.1

Drain the water tank via the menu

Pump water into the chamber
The DAC UNIVERSAL is switched on and ready for operation.
The water supply is disconnected (for a permanent water supply).
The STANDARD lid of the DAC UNIVERSAL is removed and there is
an unobstructed view into the chamber.
6 Cleaning and maintanance DAC UNIVERSAL ADVANCED
6.5 Drain the water tank of the DAC UNIVERSAL ADVANCED