Ssb Phase Noise Sensitivity Test (For Standard); Ssb Phase Noise Sensitivity Test (For Option 011); Sa Relative Level Accuracy Test - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

Offset
Test limit [dB]
Frequency [Hz]
± 3.0
- 100
± 3.0
100
± 2.0
1 k
± 2.0
1 M

SSB Phase Noise Sensitivity Test (for Standard)

CW Frequency: 70 MHz, Correlation = 1, Start frequency = 1 Hz
Offset Frequency

SSB Phase Noise Sensitivity Test (for Option 011)

CW Frequency: 70 MHz, Start frequency = 10 Hz
Offset Frequency

SA Relative Level Accuracy Test

CW Frequency: 100 MHz
Offset Frequency
Test limit [dB]
[Hz]
± 1.5
-7.5 M
± 1.5
-6.5 M
± 1.5
-5.5 M
± 1.5
-4.5 M
± 1.5
-3.5 M
Chapter 2
Test result
[dB]
Test limit
[Hz]
[dBc/Hz]
1 M
< -168
3 M
< -169
7 M
< -170
Test limit
[Hz]
[dBc/Hz]
1 M
< -156
3 M
< -157
7 M
< -159
Test result
[dB]
E5052B Performance Test Record
Measurement
TAR
uncertainty [dB]
± 0.252
16.63
± 0.253
16.62
± 0.251
9.81
± 0.223
11.09
Test result
Measurement
[dBc/Hz]
uncertainty
[dBc/Hz]
± 3.13
± 2.83
± 2.57
Test result
Measurement
[dBc/Hz]
uncertainty
[dBc/Hz]
± 3.02
± 3.24
± 2.82
Measurement
TAR
uncertainty [dB]
± 0.279
6.22
± 0.292
5.94
± 0.304
5.69
± 0.308
5.63
± 0.325
5.31
Performance Tests
Status
PASS
PASS
PASS
PASS
Status
PASS
PASS
PASS
Status
PASS
PASS
PASS
Status
PASS
PASS
PASS
PASS
PASS
37