Command Reference
E4981A SCPI Commands

[:SENSe]:AVERage[:STATe]

Syntax
[:SENSe]:AVERage[:STATe] {ON|OFF|1|0}
[:SENSe]:AVERage[:STATe]?
Description
Turns ON/OFF the averaging function.
Parameters
ON or 1 (initial value)
OFF or 0
Query response
{1|0}<newline><^END>
Related commands
[:SENSe]:AVERage:COUNt
Equivalent key
[Meas Setup] - AVG - ON/OFF
sequence
Setting the averaging count with the front panel keys will automatically
turn on the function.

[:SENSe]:CORRection:CKIT:STAN1:FORMat

Syntax
[:SENSe]:CORRection:CKIT:STAN1:FORMat {GB|CPG}
[:SENSe]:CORRection:CKIT:STAN1:FORMat?
Description
Sets the parameter types of the OPEN correction data.
Parameters
GB (initial value)
CPG
Query response
{GB|CPG}<newline><^END>
Related commands
[:SENSe]:CORRection:DATA
Equivalent key
[Meas Setup] - CORRECTION - OPEN - G-B/Cp-G
sequence
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
Description
Turns ON the averaging function.
Turns OFF the averaging function.
on page 232
Description
Specifies G as the primary parameter and B as the secondary
parameter.
Specifies Cp as the primary parameter and G as the secondary
parameter.
on page 240