[:Sense]:Average:count - Keysight E4981A Programming Manual

Capacitance meter
Hide thumbs Also See for E4981A:
Table of Contents

Advertisement

Command Reference
E4981A SCPI Commands

[:SENSe]:AVERage:COUNt

Syntax
[:SENSe]:AVERage:COUNt <numeric>
[:SENSe]:AVERage:COUNt?
Description
Sets the averaging count of the measured value for the averaging function.
Unlike setting the averaging count with the front panel keys, using this
command to set the averaging count does not automatically turn on the
averaging function. Therefore, if the averaging function has been set to off, you
have to turn it on by using the
Parameters
Description
Range
Initial value
Resolution
If the specified parameter is out of the allowable setup range, the minimum
value (if the lower limit of the range is not reached) or the maximum value (if
the upper limit of the range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Related commands
[:SENSe]:AVERage[:STATe]
Equivalent key
[Meas Setup] - AVG - INCR + / DECR-
sequence
232
[:SENSe]:AVERage[:STATe]
<numeric>
Averaging count
1 to 256
1
1
on page 233
Keysight E4981A 120 Hz/1 kHz/1 MHz Capacitance Meter
command.