System:preset; System:time; System:error[:Next]; System:error:count - Keysight N774 C Series Programming Manual

Multiport power meter
Table of Contents

Advertisement

3
Instrument Setup and Status
Parameters:
24-hour time format: hours (0-23), minutes (0-59), seconds (0-59)
Response:
none
Example:
:syst:time 20,15,30
Command:

:SYSTem:TIME?

Syntax:
:SYSTem:TIME?
Description:
Returns the instrument's internal time.
Parameters:
none
Response:
The time in the format hour, minute, second. Hours are counted 0...23 (24 hour time format).
Example:
:syst:time?
Command:

:SYSTem:ERRor[:NEXT]?

Syntax:
:SYSTem:ERRor[:NEXT]?
Description:
Returns the next error from the error queue.
Parameters:
none
Response:
The number of the latest error, and its meaning.
Note: Every connection uses its own error queue
Example:
:syst:err? -> -113,"Undefined header"
Command:

:SYSTem:ERRor:COUNt?

Syntax:
:SYSTem:ERRor:COUNt?
Description:
Returns the total no. of errors.
Parameters:
none
Response:
The total count of errors.
Example:
:syst:err:coun? -> 20
44
->
+20,+15,+30