Figure 1-4
Controlling Keysight B2980
Using the Trace Buffer
Using the Trace Buffer
This section describes how to use the trace buffer.
•
"Setting the Trace Buffer"
•
"Reading the Trace Data"
Trace Buffer and Data Flow
FETCh?
READ?
MEASure?
SENS:DATA?
CALC:MATH:DATA?
CALC:DATA?
1- 20
Sense
(Current, Charge, Voltage)
VOLT
CURR
CHAR
RES
TIME
STAT
TEMP
SOUR
HUM
MATH
CALC
TIME
STAT
Trace
Feeding
VOLT
CURR
CHAR
RES
CALC
TIME
Binning
STAT
TEMP
SOUR
HUM
TRACe:DATA?
TRACe
:BIN
Keysight B2980 Programming Guide, Ed ition 1
Feeding
Limit Test
Offset
Binning
CALC
TIME
STAT
MEAN
SDEV
MIN
MAX
PKPK
TRACe:STAT:DATA?
:DATA?