MEMory[:METer][1]:TABLe:GAIN[:MAGNitude]:POINts?
Returns the number of offset points for the currently selected table.
MEMory[:METer][1]:TABLe:GAIN[:MAGNitude]?
Returns a list of offsets for the currently selected table.
Refer to the individual command under
on page 315 for details.
Modifying table data
If you need to modify the frequency and offset data stored in a
frequency-dependent offset table, you need to resend the complete data lists.
If you have retained the original data in a program, edit the program and resend
the data.

Selecting frequency-dependent offset tables

After you have edited a frequency-dependent offset table, you can select it using
the following command:
[SENSe[1]:]CORRection:CSET2[:SELect] <string>
To determine which frequency-dependent offset table is currently selected, use
the following query:
[SENSe[1]:]CORRection:CSET2[:SELect]?
Refer to
details.

Enabling a frequency-dependent offset table

To enable a frequency-dependent offset table, use the following command:
[SENSe[1]:]CORRection:CSET2:STATe ON
If you enable this command and no frequency-dependent offset table has been
selected, error –221,"Settings conflict" occurs and this command remains
disabled.
Refer to
details.
80
"[SENSe[1]:]CORRection:CSET2[:SELect] <string>"
"[SENSe[1]:]CORRection:CSET2:STATe <boolean>"
"MEMory[:METer][1]:TABLe Commands"
on page 360 for
on page 364 for
Keysight N432A Programming Guide