Modifying And Adding Calibration Factor Tables - Keysight 53147A Operating And Programming Manual

Microwave frequency counter/power meter/dvm
Table of Contents

Advertisement

11 Disconnect the power-head input connector from the Power Meter Output
12 To measure the signal power in Watts, press the Shift key, and then press

Modifying and Adding Calibration Factor Tables

The HEAD menu option provides access to preconfigured, editable
calibration-factor tables for five models of Keysight power-sensor heads (models
8481A, 8481D, 8482A, 8485A, and 8487A) and three custom tables. You can
modify the frequency/calibration-factor values in any of the data points for any
power head, and you can input data to build new calibration tables (CUST 1,
CUST 2, and CUST 3).
the various data points in the five factory-configured calibration-factor tables (for
models 8481A, 8481D, 8482A, 8485A, and 8487A) and the three custom
calibration tables (CUST 1, CUST 2, and CUST 3).
Keysight 53147A/148A/149A Operating and Programming Guide
connector and connect it to the signal to be measured.
The Power Meter measures the signal power and displays it, as shown below:
Ch 2
Freq
Pwr
the dBm/W (Display Power) key.
When you press the Shift key, the Shift annunciator is activated. When you
press the dBm/W (Display Power) key, the Shift annunciator goes off, and the
units of measurement annunciator group to the right of the digital power
measurement changes from dB or dBm to Watts, mW, μW, or nW, as shown
below:
Ch 2
Freq
Rel
Pwr
Table 2-1
Operating Your Instrument
dB
dBm
Watts
mW
uW
%
dB
dBm
Watts
mW
uW
%
shows the changes you can and cannot make to
2
97