5 Press
6 To confirm your choice, press the appropriate unit softkey.
7 Connect the power sensor to the signal to be measured.
8 The corrected measurement result is now displayed.
If the measurement frequency does not correspond directly to a frequency in the
NOTE
sensor calibration table, the power meter calculates the calibration factor using
linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor
calibration table, the power meter uses the highest or lowest frequency point in
the sensor calibration table to set the calibration factor.
When
NOTE
the sensor table identifier is displayed in the upper window. Also, pressing
each channel derived from the selected sensor tables.
Figure 5-8
Keysight N1913/1914A User's Guide
to display the Frequency pop-up window. Use the numeric
keypad to enter the required value in the Frequency pop-up window.
Single Numeric
Offset
,
displays the frequency you entered and calibration factor for
Frequency/calibration table display
Using 8480 Series Power Sensors
display mode is chosen, the frequency you entered and
5
165