Tests and Adjustments
Adjustments
Procedure
1. Connect the equipment as shown in
between the network analyzer and the power meter.
Figure 3-16
Equipment Setup for the Source Adjustment
2. .Press UTILITY
click Source Adjustment.
3. Ensure the GPIB settings are correct.
4. Follow the instructions and prompts as they are displayed.
IF Gain Adjustment
The IF gain adjustment is used to adjust the IF gain of the network analyzer.
Procedure
1. Press UTILITY
click IF Gain Adjustment.
2. Follow the instructions and prompts as they are displayed.
Receiver Characterization
This characterizes the receivers in your analyzer.
Procedure
1. Press UTILITY
click Receiver Characterization.
Keysight N5224B/25B Service Guide
Figure
Service
System
, then
Service
System
, then
Service
System
, then
3-16. Connect a GPIB cable
Adjustment Routines
, then
Adjustment Routines
, then
Adjustment Routines
, then
, then
, then
, then
3-47