3 Installation: Panoramic X-ray unit

3.4 Installing the floor stand (option 2)

3.4 Installing the floor stand (option 2)
1.
2.
D
CAUTION
For installation with the floor stand, the unit remains lying on
the pallet until the floor stand has been completely assem -
bled. Only then may the unit be installed. For enhanced rep -
resentation, some of the following drawings are shown in
the standing state.
1. Remove the surrounding packaging, the two lateral sty -
rofoam parts and profile cover A.
NOTICE
The center styrofoam part should remain attached to the
unit for protection.
2. Loosen the nuts B (with washers) on both sides of the
pallet (or the wooden support). Take off the pallet (or
the wooden support). Remove the threaded bolt C.
30
A
Must be dis -
mantled ly -
ing on its
C
C
C
B
B
Installation Instructions ORTHOPHOS XG 3
3.
E
F
4.
G
F
IMPORTANT
The nuts D on the unit may remain inside the unit when the
threaded rods are removed. Remove the nuts.
3.
Remove the carrying handles E on the lower side of the
pallet from the holders F.
4.
Loosen screws G and remove the holders F.
Sirona Dental Systems GmbH