Location 12: Can Bus; Location 13: Stand/Peripherals - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
Error code
Description
E6 12 01
CAN controller initialization error on
DX1.
Error code
Description
E6 12 02
CAN malfunction (cannot be
assigned to module).
Error code
Description
E4 13 01
Actuator 1 has not reached pan
home position.
Error code
Description
E4 13 02
Actuator 1 has not left pan home
position.
63 03 510 D3352
D3352.076.03.11.02 09.2016
5.5.5

Location 12: CAN bus

Actions required
E6 12 01
● Check the CAN bus.
Actions required
E6 12 02
● Check the CAN bus.
5.5.6

Location 13: Stand/Peripherals

Actions required
E4 13 01 XG 3D
● Check the actuator mechanisms manually
for smooth and easy running.
In case of stiffness ...
● manually reset the actuators to the zero
position and run a test cycle via P1.
If the error occurs again ...
● Check light barrier V1_1 (X802) with
service routine S015.5, replace light
barrier if necessary.
● Run service routine S015.5 (free travel) to
check electronic activation and jerk-free
running, replace actuators and/or spindles
and board DX1 if necessary.
Actions required
E4 13 02 XG 3D
● Check the actuator mechanisms manually
for smooth and easy running.
In case of stiffness ...
● manually reset the actuators to the zero
position and run a test cycle via P1.
If the error occurs again ...
● Check light barrier V1_1 (X802) with
service routine S015.5, replace light
barrier if necessary.
● Run service routine S015.5 (free travel) to
check electronic activation and jerk-free
running, replace actuators and/or spindles
and board DX1 if necessary.