(Only For Xg Plus Full Version / 3D Ready ); S017: Test Step 13 - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

S017.13: Enabling/disabling the welcome screen
5.13 Service routine S017
Enable/disable the welcome screen
5 – 104
5.13.11

S017: Test step 13

(only for XG
Start test step 13 in the same way as described on page 5-85 or by press-
ing the Service key
i
NOTE
If you go to test step 13 from test step 12 by pressing the Service key, the se-
curity access confirmation will be skipped.
1.
Select the code for enabling or disabling the welcome screen using the ar-
row keys of selection field 1.
00 = Welcome screen disabled
01 = Welcome screen enabled
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
-
01
2
3
R
2.
After the code is selected the memory key
lection, first press the memory key
R key
.
R
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
-
01
2
3
R
The welcome screen is enabled or disabled.
i
NOTE
After the welcome screen is disabled all parameters of the service routine
S017.14 are reset back to the factory setting.
Plus
ready
full version / 3D
in test step 12.
+
13
lights up. To save the se-
(the R key lights up) and then the
+
S17
13
D3352.076.01.26.02 09.2016
Tab 5
)
59 38 399 D3352

Hide quick links:

Advertisement