Pre-Programmed Exposure Times For Films Of Sensitivity Class E And With A 300 Mm (12") Fha Cone - Sirona Heliodent Plus Operating Instructions Manual

X-ray system
Hide thumbs Also See for Heliodent Plus:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions HELIODENT
Upper
jaw
Lower
jaw
Upper
jaw
Lower
jaw
Exposure time in
seconds with:
60 kV
70 kV
Freely programmed
values
62 14 717 D3507
D3507.201.01.06.02 05.2011
PLUS
4.6.3
Pre-programmed exposure times for films of sensitivity
class E and with a 300 mm (12") FHA cone

Pre-programmed exposure times for films of sensitivity class E and with a 300 mm (12") FHA cone

0.12
0.16
0.20
0.06
0.08
0.10
NOTICE
For film of sensitivity class F: Set the exposure time one level lower with
the minus button.
For films of sensitivity class D: Set the exposure time four levels higher
with the plus button.
Using a film holder: Set the exposure time one or two levels higher with
the plus button.
4 Controls and functional elements
0.25
0.32
0.40
0.12
0.16
0.20
4.6 Exposure times
0.50
0.64
0.25
0.32
19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents