S017: Test step 12
11.10.3.8
Enabling/disabling the preview image
IMPORTANT
The preview image is automatically disabled for volume exposures,
regardless of the setting in service routine S017.12.
Selection field
01
* Factory setting
1. Call service routine S017.12 [ → 125].
 Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use arrow keys (A) to select the required setting in selection field 1
(see table).
 Once the required setting has been selected, the Memory key
lights up.
3. Save the setting [ → 129].
4. Exit the service routine [ → 129].
11.10 Service routines for startup
Code
Function
00
Preview image
disabled
01
Preview image
enabled*
11 Initial startup