Malfunctions; Error Messages - Sirona teneo Operating Instructions Manual

Hide thumbs Also See for teneo:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions (USA) TENEO
Code
Error
10
Flow rate of the SPRAYVIT
M on dentist element too
low
11
Flow rate of SPRAYVIT M
on assistant element too
low
12
Tumbler temperature fuse
defective
13
Battery of wireless foot
control empty
14
Consumption of waterline
treatment solution too low
15
Water feed too low
16
Hydrocolloid flow rate
17
Emergency pump operation Malfunctioning of pump sensors in the
18
Chair movement outside of
the permissible movement
limits or chair drive position
not plausible
65 48 510 D3509
D3509.201.03.03.23 10.2016
7

Malfunctions

7.1

Error messages

Any error states of the dental treatment unit that cannot be immediately
recognized by the user and must nevertheless be corrected are displayed
on the touchscreen. If an error state exists, this is indicated by a warning
triangle A in the status line of the touchscreen.
➢ Touch the warning triangle A in the status bar of the touchscreen.
 The error code(s) is(/are) displayed.
The error codes have the following meaning:
Description
Error table, without HF surgery
The water flow rate is below the limit.
The heater of the SPRAYVIT M is no
longer heated to prevent scalding by
overheated water.
The temperature fuse was blown by
overheating. The water in the tumbler
filling unit is no longer heated.
Operation of the dental treatment unit
no longer possible due to fully
depleted battery.
The tank for waterline treatment
solution was not emptied since its last
filling even though many disinfection
cycles were performed.
The maximum filling time for the
mixing tank is exceeded.
It was detected via the water pump
that the flow rate of the hydrocolloid is
too low.
water unit.
If the chair is located outside of the
permissible movement limits or the
position is not plausible, the chair
stops moving.
7.1 Error messages
Measure
Clean the outlet nozzle of the
SPRAYVIT M, see the "SPRAYVIT M
operating instructions and perform
the flow test, see "Checking the flow
rate on the SPRAYVIT M
multifunctional syringe".
Inform your service technician.
See "Changing the battery of the
wireless foot control" [ → 245].
Inform your service technician
Change the water filter, see
"Changing the water and air filters"
[ → 236].
Check the water pressure of the
drinking water supply.
Check the hydrocolloid hoses for
clogging.
Inform your service technician.
Inform your service technician.
7 Malfunctions
263