Table of Contents

Advertisement

5 Messages
88
5

Messages

The different message texts are displayed ...
● On the Easypad touchscreen
XG 3D message texts
● On the display of the remote control
There are 3 groups of message texts:
Help messages (Hx xx):
● Help messages are caused by operator errors.
● The user must take action.
Error messages (Ex yyxx):
● Error messages indicate unit faults.
● The user must take action to eliminate the fault(s).
System messages (Sxxx):
● System messages inform the user about the current operating status
of the unit.
● The user does not have to take any action.
If error messages are displayed on the control panel that are not listed in
this section (such as message 1311), these messages come from the
Windows system. In such cases, you must check whether the firmware
you are using is compatible with the SIDEXIS XG version and run a
software update [ → 62] if necessary.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016