9 Repair
9.16 Boards
Board
DX1
DX11
DX32
432
9.16.3

Measures following replacement of boards

IMPORTANT
Measures following replacement of boards: XG 3D
After replacing boards or modules containing boards, check to make
sure that the software version of the module corresponds to the current
software version of the system.
The software versions of the modules can be queried by running service
routine S008.2 [ → 286] or using the extended detail query in SiXABCon.
You can also check the info screen in advance to determine whether the
current software constellation is permissible. If this is not the case, the
version number of the entire software is marked by an asterisk (e.g.
V04.03.01*).
In case of software incompatibilities, carry out a software update or
downgrade [ → 62].
Always carry out the measures described below exactly in the order
specified and do not perform any other actions in between.
The following table provides an overview of various possible replacement
situations and cross-references to detailed descriptions of the actions
required for the corresponding situations following board replacement:
Constellation
Inserting a new DX1
Unit software version V04.04.00 or
higher
Inserting a new DX11
IMPORTANT: Observe the current information concerning the software compatibility
with SIDEXIS XG (see Release Notes on the accompanying CD or on the Internet page
of the SIRONA dealer).
IMPORTANT: Use the most current software version available on the ORTHOPHOS
XG3D CD.
Unit software version V04.04.00 or
higher
Inserting a new DX32
Unit software version V04.04.00 or
higher
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
Measures
● Switch the unit on.
● Perform a complete unit
adjustment or calibration.
For units with an occlusal bite block:
● Perform adjustment of the AD
acceptance angle values using
service routine S018.9.
● Proceed exactly as described in
the section titled "".
● No further action is required.
D3352.076.03.11.02 09.2016
see
S. [ → 184]
S. [ → 334]
S. [ → 435]