8 Maintenance and service
8.6 Replacement of parts subject to wear and tear
68
CAUTION
Make sure that the small metal cylinder of the cable is properly placed
in the anti-pull protection. Fiber in cable may break if not correctly
mounted resulting in high repair costs.
Only use the Sirona Dental Systems battery pack, see "Spare parts"
[→ 23].
Replacement of parts subject to wear and tear
8.6
Check the following parts subject to wear and tear and replace where
applicable:
● Optical fibers (change after 10 treatments/sterilizations)
● Single-use tips (change after every treatment)
● Handpiece sleeve incl. plastic body and silicone switch coat (change
after 400 treatments/sterilizations)
● Bending tool for single-use tips (change after 100 treatments/
sterilizations)
● Fiber cutter (change after 400 treatments/sterilizations or every two
years)
● Rechargeable battery (change after 1000 charging cycles or every
two years)
For further informations, see chapter "Cleaning, disinfection and
sterilization" [→ 61].
CAUTION
Only use parts from Sirona Dental Systems, see "Spare parts" [→ 23].