Replacing board DX11

9.16.2.2
IMPORTANT
Replacing DX11 information: GALILILEOS is old
The software version of the "DX1/DX11 board" must be compatible with
the main software version of the unit [ → 49].
IMPORTANT
The cover plates of the DX1 (REF 59 24 142 and REF 62 82 052) board
versions are not compatible with each other.
9.16.2.2.1
In the case of boards with REF 59 24 142 (DX1) and REF 59 25 214
(DX11)
The cover plate of board DX11 must be removed [ → 430].
1. CAUTION! Touch a ground point to discharge static electricity before
touching any boards.
Pull the defective DX11 board to remove it from the DX1 board.
2. Insert the new DX11 board on the DX1 and reattach the cover plate.
9.16.2.2.2
In the case of boards with REF 62 82 052 (DX1) and REF 63 17 056
(DX11)
The cover plate of board DX11 must be removed [ → 430].
1. CAUTION! Touch a ground point to discharge static electricity before
touching any boards.
Pull the defective DX11 board using the removal tool (D) to remove it
from the DX1 board.
IMPORTANT: The removal tool (D) is included in the delivery scope
of the DX11 board.
2. NOTICE! Ensure that the connector strips of boards DX1 and DX11
are aligned precisely above one another and are not offset, before
D
pressing the boards together firmly.
Attach the new DX11 board on to the DX1.
3. Check that the DX11 board is correctly attached to the DX1 board.
DX11
There must be no visible gap between the connector strips.
4. Reattach the cover plate.
After replacing the board DX11, you must observe the notes provided in
Replacing DX11: XG 3D
the chapter Measures following replacement of boards [ → 432] .
9.16.2.3

Replacing board DX32

The removal of the board DX32 is described in the chapter Removing
board DX32 [ → 374] . Install the board by following the same procedure
in reverse order.
After replacing the board DX32, you must observe the notes provided in
Replacing DX32: XG 3D
the chapter Measures following replacement of boards [ → 432] .