6.3

Replacing the set screw

NOTICE
Not to be confused
Do not confuse the set screw with the ball pressure screw!
NOTICE
Wear to the set screw
Replace the set screw every 1000 clamping procedures or if the pin has
broken off.
1. If a block is inserted in the block fixing, remove it.
2. In the software, navigate to the system menu, and click on the
"Configuration"
3. Click on the
4. Click on the unit whose set screw you wish to replace.
5. Click on the step
 The milling and grinding unit moves into position (12 o'clock
position) to insert the set screw.
 A dialog box prompts you to replace the set screw and close the
processing chamber door again.
6. Remove the worn set screw using the block clamp tool.
7. Take a new set screw out of the drawer of the unit, insert it, and
secure it in place using the block clamp tool.
8. Click on the
button.
"Devices"
button.
"Change set screw"
"Ok"
button.