Aiming beam
Interlock
Overheating
Accustic signal
63 74 198 D3545
D3545.201.06.03.02 05.2014
There is no aiming beam.
● Check to see if the optical fiber or its connector is damaged. If the
optical fiber is damaged, replace it with a new one.
● Check the connection of the optical fiber.
● Check the proper assembly of the handpiece sleeve.
● Be sure that all operational steps have been carried out correctly.
The light of the aiming beam does not project a uniform circular pattern.
● Trim the end of the optical fiber with the fiber cutter again. Always
make the notch perpendicular to the optical fiber.
The interlock is claimed to be open.
Interlock is used:
● Check the connection of the interlock.
● Check if the door is open.
Interlock is not being used:
● Check if the interlock bridge is connected properly.
The laser source is claimed to overheat.
● Check if all convection openings for air cooling on the sides of the unit
are uncovered.
● Check if unit is placed near heat sources. If so, place the unit
somewhere else and let it cool down.
There is no accustic signal for activating the laser and/or pushing the
buttons.
● Check the settings for the accustic signals in the set-up submenu.
Cannot solve the malfunctioning, Sirona
If you cannot solve the malfunctioning, switch off the laser and contact
Sirona Dental Systems GmbH, your local dental depot or your authorized
service center.
8.4

Technical support, repair and testing

Technical information on the repair, Sirona
Sirona provides technical information on the repair of individual
components only to authorized dealers and only after conducting an
advanced training course for their technical personnel. Please contact
your local dental depot or an authorized Customer Service Department
for technical support.
The SIROLaser Advance may be sent in for repair or for safety inspection
only in its original packaging, including all accessories. Disinfect the
SIROLaser Advance and sterilize the accessories according to the
relevant instructions for use before shipping them.