8 Maintenance and service

8.4 Technical support, repair and testing

Interlock
Overheating
Accustic signal
66
The interlock is claimed to be open.
Interlock is used:
● Check the connection of the interlock.
● Check if the door is open.
Interlock is not being used:
● Check if the interlock bridge is connected properly.
The laser source is claimed to overheat.
● Check if all convection openings for air cooling on the sides of the
unit are uncovered.
● Check if unit is placed near heat sources. If so, place the unit
somewhere else and let it cool down.
There is no accustic signal for activating the laser and/or pushing the
buttons.
● Check the settings for the accustic signals in the set-up submenu.
If you cannot solve the malfunctioning, switch off the laser and contact
Sirona Dental Systems GmbH, your local dental depot or your
authorized service center.
8.4
Technical support, repair and testing
Sirona provides technical information on the repair of individual
components only to authorized dealers and only after conducting an
advanced training course for their technical personnel. Please contact
your local dental depot or an authorized Customer Service Department
for technical support.
The SIROLaser Xtend may be sent in for repair or for safety inspection
only in its original packaging, including all accessories. Disinfect the
SIROLaser Xtend and sterilize the accessories according to the relevant
instructions for use before shipping them.