11 Initial startup
11.10 Service routines for startup
SR*
S018
S018.2
S018.3
S018.4
S018.5
S018.6
S018.7
S018.8
S018.9
S018.10
152
11.10.4

S018: Service for height adjustment

SHZ**
Function
S018 XG 3D / XG 3D ready
Service for height adjustment
No
Setting the maximum travel height
No
Undoing the maximum travel height setting
No
Checking the height adjustment sensor system
No
Setting the minimum travel height
No
Undoing the minimum travel height setting
No
Setting the nominal angle for the occlusal bite block (as required by the
customer) (occlusal bite block only)
NOTICE! This routine may only be selected for units with occlusal bite
block; otherwise, the unit will malfunction!
No
Function test of the occlusal bite block (occlusal bite block only)
NOTICE! This routine may only be selected for units with occlusal bite
block; otherwise, the unit will malfunction!
No
Adjustment of the AD acceptance angle values and the zero point
NOTICE! This routine may only be selected for units with occlusal bite
block; otherwise, the unit will malfunction!
No
Setting the PWM values for the height adjustment creep speed
(PWM=Pulse Width Modulation)
NOTICE! This routine may only be selected for units with occlusal bite
block; otherwise, the unit will malfunction!