Tab 5
Function S015
Actuator service
Test step 5
Testing of actuators 1 and 2
Selecting service routine S015
Selecting test step 5
59 38 399 D3352
D3352.076.01.26.02 09.2016
5.12
Service routine S015
Actuator service
Select the Service menu (see page 5-12).
1.
Use the arrow keys to select service routine S015.
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
+
-
S015
+
-
5
2
3
5.12.1
S015: Test step 5
The use of this service routine is suitable for troubleshooting motor and cable
defects. Furthermore, the possibility of performing isolated travel with an indi-
vidual drive makes it possible to release the clamping of the actuator/rotation
system and check the stiffness of the individual adjusting axes.
1.
Select test step 5 in selection field 2 with the arrow keys and confirm your
selection by pressing the Service key
Plus
ready
XG
/ 3D
:
Easypad touchscreen
1
+
-
S015
-
5
+
2
3
S015: Actuator service
5.12 Service routine S015
XG 5 / 3 / 3 PPE: Multipad
➊
➊
S015_______5____
1
2
.
XG 5 / 3 / 3 PPE: Multipad
S015_______5____
➊
1
2
3
➊
3
5 – 81