S012: Can Bus Service; S012: Test Step 1 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

8 Perform service routines via the control panel
8.11 S012: CAN bus service
SR*
S012
S012.1
S012.2
S012.3
S012.4
Checking the module
294
8.11

S012: CAN bus service

SHZ**
Function
S012 XG 3D
CAN bus service
No
Presence display of modules
No
Inquiry of the CAN status register of the modules
No
Resetting the CAN status register of the modules
No
Display of CAN bus cycle on LEDs of modules
* SR=service routine, ** SHZ=security access
IMPORTANT
The CAN bus service is not yet implemented for the module DX11!
8.11.1

S012: Test step 1

Presence display of modules
Selection field
1
2
➢ Call service routine S012.1.
S012.1: XG 3D
➢ Use the arrow keys (A) in selection field 1 to select the required
module.
Service Manual ORTHOPHOS XG 3D / Ceph
Parameter/Display
Subassembly
● Counter value of
CAN bus events
● Presence code behind
the counter value:
P = module present
L = module lost
The counter value of the CAN bus events (change from "P" to "L"
since the last switch-on of the unit) of the selected module
processed so far is displayed in selection field 2 with the
presence code of the module ("L" or "P") (see table).
Once the module has been selected, the T key (B) lights up.
Sirona Dental Systems GmbH
Range of
values
DX1 - DX88
63 03 510 D3352
D3352.076.03.11.02 09.2016

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents

Save PDF