S011: Dosimetry (Without Ring Movement); S011: Test Step 14 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
SR*
S011
S011.14
S011.24
63 03 510 D3352
D3352.076.03.11.02 09.2016
8.10

S011: Dosimetry (without ring movement)

Unit is radiating X-rays
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
SA**
Function
S011: XG 3D
Dosimetry (without ring movement)
yes
Dosimetry with continuous radiation
yes
Dosimetry with pulsed radiation
* SR=service routine, ** SA=security access
8.10.1

S011: Test step 14

Dosimetry with continuous radiation
A standard dosimeter is required for dosimetry.
Selection field
S14.11 XG3D
1
2
1. Attach the dosimeter sensor in the middle of the combisensor.
2. Call service routine S011.14 [ → 258].
3. Use the arrow keys (A) to select the required kV/mA level, e.g. "85kV/
5mA".
4. Save the setting [ → 262].
8 Perform service routines via the control panel
8.10 S011: Dosimetry (without ring movement)
WARNING
Parameters
kV/mA levels
Radiation time
(not selectable)
The kV/mA level is displayed in selection field 1.
The corresponding radiation time is displayed in selection field 2.
The corresponding radiation time "3.2 s" is displayed in selection
field 2.
NOTE: The radiation time cannot be selected separately.
Once the required setting has been selected, the Memory key (B)
lights up.
Range of values
85kV/5mA (3.2s)
85kV/7mA (5s)
85kV/10mA (5s)
85kV/13mA (5s)
3.2 s
5 s
291