6 Remote Control
6
Remote Control
i
NOTE
If a door contact is required, connect to K10 according to
Wiring References.
6.1
First installation possibility: Multitimer without spiral cable
ATTENTION
The Multitimer with spiral cable is temporarily connected to
the unit's carriage for start-up.
•
Therefore the installation described on the following
page cannot be performed until after the "Phantom Ra-
diograph" section.
•
After installing the Multitimer without spiral cable, an-
other functional check is required:
Take and evaluate a phantom radiograph.
12
ATTENTION
Keep the spiral cable left over after installation at the cus-
tomer's for later service use.