Remote Control; First Installation Possibility: Multitimer Without Spiral Cable - Sirona orthophos plus Installation Instructions Manual

Table of Contents

Advertisement

6 Remote Control

6
Remote Control
i
NOTE
If a door contact is required, connect to K10 according to
Wiring References.
6.1

First installation possibility: Multitimer without spiral cable

ATTENTION
The Multitimer with spiral cable is temporarily connected to
the unit's carriage for start-up.
Therefore the installation described on the following
page cannot be performed until after the "Phantom Ra-
diograph" section.
After installing the Multitimer without spiral cable, an-
other functional check is required:
Take and evaluate a phantom radiograph.
12
ATTENTION
Keep the spiral cable left over after installation at the cus-
tomer's for later service use.
Sirona Dental Systems GmbH
60 82 064 D 3297
D 3297.031.01.04.02

Advertisement

Table of Contents
loading

Table of Contents