13 Adjusting and calibrating the unit
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
220

Pan symmetry

13.2.1.3
13.2.1.3.1
Automatic adjustment: Pan symmetry
The OP XG needle phantom is in the pan bite block holder on the unit.
1. In the structure tree, under
Symmetrie"
 The
2. Make SIDEXIS XG ready for exposure. [ → 206]
3. Take an exposure (60 kV/3 mA; 14.1 s) [ → 206].
 SIDEXIS XG calculates the adjustment values S1, S2, and S3
automatically from the exposure and enters them in the text
boxes of the
 The exposure is displayed in the exposure window in SIDEXIS
XG.
Installation Manual ORTHOPHOS XG 3D / Ceph
"2D Adjustment"
element (S010.2).
"Pan - Symmetrie"
menu is displayed in the action area.
"Pan - Symmetrie"
Sirona Dental Systems GmbH
"Pan -
, click on the
menu.
63 03 452 D3352
D3352.031.05.08.02 10.2015