Operation DAC UNIVERSAL ADVANCED

5.1

Regular activities

The DAC UNIVERSAL was installed by a service technician according to
Introduction regular activities
the instructions in the "Installation [ → 49]" section.
See below for an overview of the activities that must be conducted during
the regular use of the DAC UNIVERSAL.
Activities and relevant chapter in the Operating Instructions
„Pre- and post-reprocessing of instruments [ → 59]"

„Checking the reprocessing result with an indicator / PCD test device

[ → 59]"
Replacing the lubrication concentrate, see "Adding/replacing the
lubrication concentrate [ → 66]"
„Checking oil supply to the chamber of the DAC UNIVERSAL
ADVANCED [ → 122]"
„Filling the water tank [ → 69]"
"Checking water supply to the chamber of the DAC UNIVERSAL
ADVANCED [ → 122]"
5.1.1

Pre- and post-reprocessing of instruments

IMPORTANT
Please follow the regulations and guidelines applicable in your country
in your practice.
IMPORTANT
Please follow the instrument manufacturer's instructions when handling
the instruments.
IMPORTANT
The chuck of the instruments must be lubricated regularly after a
mechanical cleaning according to the requirements of the instrument
manufacturer.
If possible let oil remain over night in the instrument and run a cycle later
in the DAC UNIVERSAL.
5.1.2
Checking the reprocessing result with an indicator / PCD
test device
IMPORTANT
Please follow the regulations and guidelines applicable in your country
in your practice.
Before using the DAC UNIVERSAL after installation or larger repairs, we
recommend checking the sterilization result using a Class 5 chemical
indicator and/or a PCD test device.