10 Appendix
62
Seal on PC slide-in module
10.3
NOTE
If the seal is broken, all warranty claims regarding the PC slide-
module automatically expire.
The PC slide-in module may be opened only by an authorized dental
technician. Only spare parts approved by us may be used in this
module.
Following a repair, the seal supplied along with the spare parts must be
affixed at the specified location (A).
PC Diagnostic Tool
10.4
10.4.1
Starting the diagnostic tool
Test
Customer Diagnostics
Technician Diagnostics
Sirona Windows Diagnostics
Tip: If one of the following steps does not work, further steps can be
found in the Service Manual.
ü The PC is switched off.
1. Switch on the PC and wait until the above message appears.
2. Hold down the "F11" key on the keyboard until the boot menu
appears.
Operating Instructions for the Acquisition Unit CEREC AC Connect
Suitable for
Suitable for the user, in order to
check the PC components.
Suitable for the technician, in order to
check the PC components.
Windows Stress Test, in order to test
the PC at permanent load.