Right Side Controls
1. Interior light control: Press to cycle light brightness (on, low, medium, high and off).
2. Interior fan on/off toggle button: Press to toggle the interior fan. The light next to this button lights up
when the fan is on.
3. Laser status light: A status light that has three signals:
On (blue light): Indicates that all the interlocks are engaged, and ready for laser operation.
o
Off: Indicates that at least one interlock is open. The system prevents the laser from firing in
o
this state.
Blinking: Indicates that one of the interlocks is not engaged. The system prevents the laser
o
from firing in this state.
4. 12 V power jack: Connects to the AC power cord. Used to power the Vevo LAZRTight.
5. Access port cover: Secures an access port that is used to run cables and tubing into the Vevo
LAZRTight. There is one access port and cover on each side of the Vevo LAZRTight. A magnetic
interlock prevents the laser from firing unless both covers are full closed.
To remove the cover, slide the cover up and out of the magnetic interlock.
Components
Vevo F2 Imaging System User Manual