Caution:
Do not use cleaning or sterilizing chemicals on the foot pedal, V evo
Animal Monitoring System, Vevo Infusion Pump or the Vevo Optical Fiber. Doing
so could cause damage.
To Clean and Disinfect an Accessory
1. Using an approved cleaner, clean the accessory of any visible gel, debris, or fluids.
2. Verify that all gel, debris and animal fluids have been removed. If necessary, repeat the previous
step with a new wipe.
3. Using an approved disinfectant, disinfect the accessory by wiping all surfaces that may have been
contaminated.
4. Ensure the accessory stays wet with the disinfectant for the recommended wet contact time. This
may require reapplication to achieve the required wet time.
5. Allow the accessory to air dry in a clean, well ventilated space. Do not dry with a cloth.
6. Inspect t he accessory, and ensure it is free of any damage, such as corrosion, discoloration, pitting,
or cracked seals. If damage is evident, discontinue use immediately, and contact FUJIFILM
VisualSonics or your local representative.
Maintenance
Vevo F2 Imaging System User Manual
451