WARNING:
expired.
Some cleaners and disinfectants can cause an allergic reaction to some
individuals.
Caution:
Do not allow cleaning solution or disinfectant into the system, laser,
transducer or accessory connectors.
Do not use strong solvents such as thinner or benzene, or abrasive cleansers,
since these will damage the exterior surfaces. Use only FUJIFILM VisualSonics
recommended cleaners or disinfectants.
List of Compatible Cleaners and Disinfectants
The materials listed in the table below are chemically compatible with the system and transducer.
Product
Tristel™ Sporicidal
Wipes
Mikrobac® Tissues
Cleanisept® wipes
CaviCide™
CaviWipesXL™
Precautions
Use only FUJIFILM VisualSonics recommended disinfectants. Using a non-recommended
l
disinfecting solution or incorrect solution strength may cause damage and void the warranty.
To avoid electric shock, always disconnect the system, laser, transducer(s) and accessories from
l
the power before cleaning or disinfecting them.
Maintenance
Ensure that cleaning and disinfecting solutions and wipes are not
System
Transducer
Vevo F2 Imaging System User Manual
Wet Contact
Laser
30 seconds
1 minute
2 minutes
3 minutes
3 minutes
Time
446